User manual

16
Connecting Pressure Cuffs
The NIBP-1010, NIBP-1020 and NIBP-1030 Blood Pressure Simulators are connected
between the blood pressure cuff and the DUT (Patient Monitor). The blood pressure cuff
should be disconnected from the DUT and a ‘T’ adapter inserted between the cuff and the
DUT. The NIBP simulator is then connected to the open side of the ‘T’ adapter as pictured
below:
NOTE: Roll the cuff tightly on itself or around a mandrel.
Connecting Patient Leads
For models with ECG output, snaps or patient lead test posts (NIBP-1030-BE only) are
provided along the sides and are identified by the markings on the overlay.
Blood
Pressure
Cuff
DUT
NIBP-
10xx
CONNECTIONS