User manual
8
MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output
Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).
The available screens can be toggled using .
PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below.
Also displayed on this screen is the peak pressure and selected output waveform
The display will resemble the following:
SCREENS
Systolic/Diastolic (mean)
Mean
Heart Rate
Respiration
Rate (Optional)
Adult 120/80
93
80
35
Adult 120/80 w/Pace
93
Pacemaker 80
35
Adult High 190/120
150
80
35
Adult Low 80/40
53
80
35
Neonatal 70/40
55
94
47
Alarm Test (Optional)
(Alternates every 30 sec)
-
47
160
00
80
Arrhythmia Seq
(NIBP-1030 Only)
-
80
35
Leak Test (NIBP-1030 Only)
-
-
-
Manometer
-
-
-
OUTPUT SELECTION
PRESSURE
ENVELOPE
OFFSET
This line is only
displayed if a non-
zero value is
programmed into
the Systolic &/or
Diastolic Shift. (See
Setup for more
information).
PRESSURE
PEAK PRESSURE
(Optional)