To run an NIBP simulation, the cuff and monitor are connected to the pressure input. Then
the measurement is initiated by the monitor and the NIBP-1010 and NIBP-1020 will output
the proper waveform based on the cuff pressure.
The NIBP output mode can be changed by pressing the Select key. Once the desired
operating mode is selected, the output will automatically begin once the correct pressure is
detected.
ECG (NIBP-1020 Only):
The ECG output is selected by pressing the UP arrow until the ECG Waveforms menu is
displayed. While in this mode, the NIBP simulation is disabled. The select key is used to
choose the desired ECG output.
ECG Alarm Mode:
This mode tests the alarms of the monitor under test. The ECG output will alternate from a
47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR with 80 brpm respiration.
The time interval between alternations is 30 seconds. While in this mode, the BP
simulation output is disabled.