The Setup Mode allows the user to adjust the configuration of the simulator. The Setup
screen can be entered using the key. The parameters can be changed by using
key to highlight the line and to toggle the available options.
The Setup screen can be exited using the key.
The following is a breakdown of the parameters available in the configuration of the unit
and their available options:
System Setup Configuration
Parameter Description Range
Systolic Shift
Adjusts the Systolic Output of the NIBP
Simulation. This is not a direct mmHg adjustment.
±50.0
Diastolic Shift
Adjusts the Diastolic Output of the NIBP
Simulation. This is not a direct mmHg adjustment.
±50.0
SpO
2
Output
This parameter determines whether the SpO
2
output pulse is active. The output drives an MSP-
2100 FingerSim Module.
Off/On
Auto Off Timer (Min)
Determines the period of inactivity before the unit is
turned OFF. A timer is started when the unit is
turned ON and is reset each time a key is pressed.
When the timer reaches the value set in this
parameter, the power is automatically turned OFF.
(NOTE: Setting this parameter to 0 disables the
Auto Off timer. When running from line power, the
unit does not automatically shut off. Auto Off timer
is inactive during a test.)
0-30 Minutes
Contrast Adjust
Sets the contrast of the display screen.
0-20
Battery Life
Available only with Battery Option installed.
Displays current life of the battery.
At 10%, a warning screen will appear.
At 0%, the unit will power down automatically.
0-100%
(Read Only)
Software
Displays current software program.
(Read Only)
SETUP