Datasheet

Absolute encoders/MAGRES
EAM360R-S - Analog
Solid shaft with synchro flange
Magnetic single- or multiturn encoders
Teach process
Activate teach process
Start teach process within 5 minutes after power on. Set teach input for
>5 seconds on HIGH and afterwards on LOW level.
DV/Status output: Oscillates after 5 seconds.
Position 1
Get encoder on position intended for min. voltage output / current output.
Set teach input for >0.1 seconds on HIGH.
DV/Status output: Switches to HIGH level for 3 seconds and ashes
shortly.
Position 2
Get encoder on position intended for max. voltage output / current output.
Set teach input for >0.1 seconds on HIGH.
DV/Status output: Switches to HIGH level for 3 seconds and ashes
shortly. If measuring range is exceeded or the limits are too close to
each other, the teaching process was not successful and has to be
repeated.
Default
Set teach input for >15 seconds on HIGH.
DV/Status output: Oscillates after 5 seconds.