User Manual

en_BA_U500_UR18_Gen2_IO-Link.docx 28/43 Baumer Electric AG
04.04.2019/ Frauenfeld, Switzerland
4.7.2 Analog Output
4.7.2.1 Output Scale
Allows to change the upper and lower values of the analog output range. For example: the default range is 0
to 10 V, this can be modified to 2 to 8V. In terms of distance values mapped to analog values the modification
would mean that the default mapping of 0V=70mm and 10 V=1000 mm, is modified to 2V = 70mm and
8V=1000 mm.
When to use: When the master requires a different voltage or current range.
Parameter name
Short Description
Rights
Unit / Allowed
values
Analog Output.AnalogMax
Set the maximum value of the analog
output
rw
Voltage: 0-10 V,
Default 10V
Current: 4-20mA,
Default: 20mA
Analog Output.AnalogMin
Set the minimum value of the analog
output
rw
Voltage: 0-10 V,
Default 0V
Current: 4-20mA,
Default: 4 mA
Figure 23 Modified analog Output range
4.7.2.2 Distance at Analog Min/Max
With this parameters the analout output characteristic is modified. By reducing the sensing range the
resolution of the analog output is improved. Inverting the analog output can be achieved by setting Analog
Max on minimum value (V) of the analog output and AnalogMin on the maximum value (V) of the analog
output
When to use: When the resolution of the analog output range should be improved.