4.7.2  Analog Output 
4.7.2.1  Output Scale 
Allows to change the upper and lower values of the analog output range. For example: the default range is 0 
to 10 V, this can be modified to 2 to 8V. In terms of distance values mapped to analog values the modification 
would mean that the default mapping of 0V=70mm and 10 V=1000 mm, is modified to 2V = 70mm and 
8V=1000 mm. 
When to use: When the master requires a different voltage or current range. 
Parameter name 
Short Description 
Rights 
Unit / Allowed 
values 
Analog Output.AnalogMax 
Set the maximum value of the analog 
output 
rw 
Voltage: 0-10 V, 
Default 10V 
Current: 4-20mA, 
Default: 20mA 
Analog Output.AnalogMin 
Set the minimum value of the analog 
output 
rw 
Voltage: 0-10 V, 
Default 0V 
Current: 4-20mA, 
Default: 4 mA 
Figure 23 Modified analog Output range  
4.7.2.2  Distance at Analog Min/Max 
With this parameters the analout output characteristic is modified. By reducing the sensing range the 
resolution of the analog output is improved. Inverting the analog output can be achieved by setting Analog 
Max on minimum value (V) of the analog output and AnalogMin on the maximum value (V) of the analog 
output 
When to use: When the resolution of the analog output range should be improved. 