4.2.4.9 Shutterline Correction (A503k Only)
Purpose: To set the shutterline correction value. See section 3.8 for more information on shutterline
correction.
Type: This is a read or write command.
Read Command: Cmd-ID R/W-Flag Data Length Data
0x98 1 3 -
Response: Cmd-ID R/W-Flag Data Length Data
0x98 0 3 3 Bytes
Write Command: Cmd-ID R/W-Flag Data Length Data
0x98 0 3 3 Bytes
Response: None
Data Format: Byte 1 Gray value to be added to the 10 bit pixel value
Byte 2 0x00
Byte 3 0x00
Data Range: The shutterline correction setting can range from 0 DN (0x000000) to 255 DN
(0x0000FF). The default setting is 12 DN (0x00000C).