Service manual

Single T-Cable foot end - White
(“T” runs along either side of Seat
Pan to attach Pendant Hand Control).
Locking
End Cap
Staff Control
Cable female end
IMPORTANT: Run a cable tie
through the mounting tube end
hole, loop the excess cable, run
the cable tie around the looped
cable, and pull tight to secure.
Cut off excess cable tie.
Footboard
with cutout
Staff Control
Assembly
(#2)
Staff Control
Bevel
Screws
(#2)
Staff Control
Bevel Screws
(#4)
(#3)
(#5)
(#1)
Shroud Screws
TROUBLESHOOTING — REPLACING THE STAFF CONTROL
!
ALWAYS UNPLUG THE POWER CORD BEFORE
PERFORMING MAINTENANCE ON THE BED.
1. Unplug the power cord from the wall outlet.
2. Unscrew the four Phillips screws (#1) that attach the
shroud (cable cover) to the footboard and set aside.
3. At the foot end of the bed, unscrew the
T-Cable retaining cap (#3) and unplug
the two cables.
4. Unscrew the two remaining Phillips
screws (#2) that attach the staff control
assembly bevel to the footboard from
the inside of the footboard. Set aside
screws for later reassembly.
5. Remove the old staff control assembly
and feed the old cable out through
the large hole.
6. Feed the cable of the new staff
control assembly through the
large hole from the outside
and insert the bevel into
the cutout on the
outside of the
footboard.
7. Secure the staff
control by reinserting
the two Phillips screws (#2)
you removed in step 4.
8. Reposition the shroud on the
inside of the footboard and
secure with the four Phillips
screws (#1) you removed in
step 2.
9. Plug the T-Cable (#5) into
the staff control cable (#4),
ensuring the phone jack plug is
seated correctly in the staff
control cable plug. Screw on the
retaining cap (#3) to secure.
Important:
10. Run a cable tie through the right footboard mounting tube large end hole.
11. Loop the excess staff control cable. Run the cable tie around the cable and pull tight to secure.
Cut off any excess cable tie.
GF Health Products, Inc. - www.grahameld.com Zenith 7000 Service Manual 999-0843-190E
ZENITH SERIES
30