Specifications

Model# F46110CM, F46118 F46120CM F46128 F46120CM-75 F46128CM-75
F46110CM-33 F46120CM,-33 F46120CM-33-75
Dimensions Overall-in Width 28-1/8 28-1/8 28-1/8 28-1/8 28-1/8 28-1/8
(cm) (71.4) (71.4) (71.4) (71.4) (71.4) (71.4)
Height 18-1/2 18-1/2 18-1/2 18-1/2 18-1/2 18-1/2
(47) (47) (47) (47) (47) (47)
Depth 19 19 19 19 19 19
(48.3 (48.3) (48.3) (48.3) (48.3 (48.3)
Chamber- Width 6 6 6 6 6 6
in (cm) (15.2) (15.2) (15.2) (15.2) (15.2) (15.2)
Height 6-1/2 6-1/2 6-1/2 6-1/2 6-1/2 6-1/2
(16.5) (16.5) (16.5) (16.5) (16.5) (16.5)
Depth 6-1/4 6-1/4 6-1/4 6-1/4 6-1/4 6-1/4
(15.9) (15.9) (15.9) (15.9) (15.9) (15.9)
Weight LBS.(kg) 152 (69.1) 152 (69.1) 152 (69.1) 152 (69.1) 152 (69.1) 152 (69.1)
Electrical Volts 240 208 240 208 240 208
Rating Amps 40*/10.4# 40*/12# 40*/10.4# 40*/12# 40*/10.4# 40*/12#
Watts 9600* 8320* 9600* 8320* 9600* 8320*
2500** 2500** 2500** 2500** 2500** 2500**
Phase 1 1 1 1 1 1
Temp. Range Cont. 1700°C 1700°C 1700°C 1700°C 1700°C 1700°C
Controller 8 Segment 8 Segment 4-Program 4-Program 20 Program 20 Program
8
General Specifications
Note
* Inrush power and current.
** Power and current required to maintain maximum temperature after stabilization.
The maximum current is determined by the limiting factor set by the current controller. In the event
that 40 amperes is not available, the current controller may be set to limit the current to some small-
er value at the expense of a somewhat longer heatup time.
The variation in current is a result of molybdenum disilicide heating elements having a large
increase in resistance with increasing temperature.
The maximum ramp rates for this furnace for heatup are: 100° per minute from 25C-1000°C and
15°C per minute from 1000°C-1700°C.