Specifications

9
3. Place the grab ring and back-up ring in the hex
nut in the order and orientation shown in Figure
3. Thread the nut onto the adapter. DO NOT use
the o-ring at this time.
4. Push the tubing through the nut until it bottoms
out in the adapter.
5. Remove the adapter nut and tubing. Place the
o-ring over the tubing. Be careful not to push the
back-up ring or grab ring further back on the
tubing when installing the o-ring.
6. Install the hex nut on the adapter and hand
tighten.
Ultrafilter and UV Lamp
Installation
Install the ultrafilter as follows:
1. Remove the pump interlock plug. Open the
cartridge access door in the rear of the unit by
sliding the latch down and pulling the door
toward you. The door will swing down. Remove
the screws securing the EASYpure UV/UF’s
cover. Remove the cover.
2. There are two “T’s” installed in the interior water
lines of the EASYpure UV/UF. One of these “T’s”
is connected to water lines which have identify-
ing labels attached to them (“Reject” and “Prod-
uct”). Remove this “T”. Ensure that the nut,
stainless-steel grab ring, backup ring and o-ring
from each connection remain on each piece of
tubing. Dispose of the “T” once you have re-
moved it.
3. Remove the ultrafilter from its protective bag.
4. Identify the feed, product and reject connectors
on the ultrafilter. The feed connector is the
solitary one on one end of the ultrafilter. Hold the
ultrafilter so the feed connectors on the top face
INSTALLATION
Figure 4: “T” Removal
Disconnect tubing from this “T” to install ultrafilter
Caution
Do not tighten tube fitting hex nut with a
wrench. Tight connections can be
easily made by hand.
Warning
Depressurize system prior to opening
cartridge access door.
Warning
A small amount of 2.0% hydrogen
peroxide is used to preserve the
ultrafilter during storage. When remov-
ing the ultrafilter from the bag, ensure
adequate ventilation and wear
protective gloves and glasses.