Specifications

6
Dimensions and Clearance Requirements
EASYpure UV/UF dimensions: 12 " W X 19" D X 17 1/8" H (30.5 cm X 48.3 cm X 43.5 cm).
Clearances:
Sides: 4" (10.2 cm) minimum to allow air flow.
Above: 12" (30.5 cm) minimum for removal of the top cover.
Cartridge replacement requires that you be able to access the back of the unit and open the cartridge access
door (total depth, unit + open door, = 34" (86.4 cm)).
Electrical Requirements
The EASYpure UV/UF is equipped with a power cord to be plugged into an electrical outlet of the appropriate
voltage.
Model D8611 - 120 VAC ±5%, -10%, 47-63 Hz.
Model D8612 - 240 VAC ±5%, -10%, 47-63 Hz.
Model D8612-33 - 230 VAC ±10%, 47-63 Hz.
Model D8613 - 100 VAC ±5%, -10%, 47-63 Hz.
Feedwater Requirements
The EASYpure UV/UF requires water pretreated by either distillation, deionization or reverse osmosis meeting
the following criteria.
TOC: Less than 1.0 ppm.
Turbidity: 1.0 N.T.U. maximum
Temperature: 40°F - 120°F (4.4°C - 48.8°C)
Resistivity (Minimum):
Distilled - 300,000 Ohms
Deionized - 1.0 Megohms
Reverse osmosis - 50,000 Ohms
Product Water
Quality:
Resistivity: ASTM Type I
TOC: Less than 2.0 ppb
Endotoxin: Less than 0.005 EU/ML
Flow rate:1.5 LPM maximum at minimum inlet feedwater pressure of 30 PSIG at 60 HZ and with a new final
filter.
General Specifications