Specifications
32
Problem Possible Causes Solutions
If a Barnstead Still is the
Short cartridge life (cont.) feedwater source, ensure that the
membrane is functioning properly.
If feedwater is from a central
water purification system, verify
water quality and proper
functioning of the system.
Water leakage inside Loose connections. Tighten connections.
EASYpure UV/UF.
Missing or defective cartridge Install or replace cartridge
O-rings. O-rings.
Green light on UV Solid State UV lamp not plugged into Plug UV lamp into UV Ballast.
Ballast not lit and audible alarm UV Solid State Ballast.
sounds.
UV lamp burnt out. Replace UV lamp.
UV lamp not connected to Recheck/reconnect.
connector or improper
connection.