Specifications
27
UV Solid State Ballast Fuse
Replacement
1. Turn off the EASYpure UV/UF and disconnect
it from the power supply. Remove the power
cord and the pump protector or interlock plug
from the rear of the unit.
2. Depressurize the system by opening the
draw-off valve, allowing water to drain until
no more flows from the valve.
3. Open the cartridge access door. Remove the
screws securing the EASYpure UV/UF cover.
4. Remove the cover by lifting straight up.
5. Locate the UV Solid State Ballast fuse,
located in the Solid State Ballast, above the
pump motor.
6. Remove the old fuse and replace with a fuse
of the same type and rating. (See Parts
Listing.)
7. Reinstall the EASYpure UV/UF cover. Close
the cartridge access door.
8. Reattach the power cord and the pump
protector or the pump interlock and reconnect
the unit to the power supply.
9. Operate normally.
Cleaning the Resistivity Cell
1. Turn off the EASYpure UV/UF and disconnect
it from the power supply. Remove the power
cord and the pump protector or interlock plug
from the rear of the unit.
MAINTENANCE AND SERVICING
Warning
Depressurize system prior to opening
cartridge access door.