Specifications
26
12. Reattach the power cord and the pump protector
or the pump interlock and reconnect the unit to
the power supply.
13. Operate normally.
Fuse Replacement
1. Turn off the EASYpure UV/UF and disconnect it
from the power supply. Remove the power cord
and the pump protector or interlock plug from the
rear of the unit.
2. Depressurize the system by opening the draw-off
valve, allowing water to drain until no more flows
from the valve.
3. Open the cartridge access door. Remove the
screws securing the EASYpure UV/UF cover.
4. Remove the cover by lifting straight up.
5. Pull out the fuse drawer located in the power
entry module.
6. Remove old fuses and replace with fuses of the
same type and rating. (See Parts Listing.)
7. Replace fuse drawer.
8. Reinstall the EASYpure UV/UF cover. Close the
cartridge access door.
9. Reattach the power cord and the pump protector
or the pump interlock and reconnect the unit to
the power supply.
10. Operate normally.
MAINTENANCE AND SERVICING
Warning
Depressurize system prior to opening
cartridge access door.
Warning
Replace fuses with those of the same
type and rating. Fuse Type 5 x 20 mm,
Time Lag (T) 250V 1.0 Amp.