Specifications

25
necessitating the EASYpure UV/UF be placed in a recircu-
lating mode for up to one hour to achieve organic content of
2 ppb TOC.
To replace the ultraviolet lamp:
1. Turn off the EASYpure UV/UF and disconnect it
from the power supply. Remove the power cord
and the pump protector or interlock plug from the
rear or the unit.
2. Depressurize the system by opening the
draw-off valve, allowing water to drain until
no more flows from the valve.
3. Open the cartridge access door. Remove the
screws securing the EASYpure UV/UF cover.
4. Remove the cover by lifting straight up.
5. Locate the UV oxidation chamber and pull the
top black plastic cover off.
Do not pull on the
cable.
6. While holding on to the lamp, remove the plug
from the lamp.
7. Remove the replacement lamp from its con-
tainer. DO NOT TOUCH THE GLASS PORTION
OF THE LAMP. It is recommended that lint free
gloves be worn when handling the lamp. The
glass portion must be free of fingerprints, perspi-
ration, etc. Even a light coating of perspiration
will reduce the effectiveness of the lamp.
8. Clean the lamp with isopropyl alcohol and a lint
free cloth.
9. Carefully insert and hold the UV lamp partially
into the UV chamber.
10. Connect the UV lamp to the receptacle in the
black plastic cover of the UV chamber. Replace
the black plastic cover on the UV chamber.
11. Reinstall the EASYpure UV/UF cover. Close the
cartridge access door.
MAINTENANCE AND SERVICING
Warning
Depressurize system prior to opening
cartridge access door.
Caution
Do not unscrew the cover, as this will
loosen the water tight seal and may
damage the replacement lamp. Pull the
cover straight up.
Note
Pay close attention to the orientation of
the plug as it is attached to the UV
lamp. It will be necessary to ensure
that the plug is installed in the same
manner as removed.