Specifications

23
8. Repeat steps 2 - 6 with the two empty car-
tridges supplied with your EASYpure UV/UF,
placing them in the center and left-hand
positions.
9. Close the cartridge access door. Remove the
0.2 micron filter and bell assembly.
10. Sanitize, install new cartridges and rinse
according to the instructions for Initial Saniti-
zation and Cartridge Installation in the Initial
Operation section.
General Cleaning Instructions
Wipe exterior surfaces with lightly dampened cloth
containing mild soap solution.
0.2 Micron Filter Replacement
Replace the 0.2 micron filter whenever any of the follow-
ing conditions occur: every 30 days, the product water
flow rate is reduced or bacteria break through. The 0.2
micron filter is shipped assembled with a bell. To replace
the 0.2 micron filter assembly:
1. Remove the old 0.2 micron filter assembly by
turning it counterclockwise until it is free from
the Luer fitting.
2. Remove the new 0.2 micron filter assembly
from its bag and insert it into the Luer fitting.
Gently turn it clockwise until it is fully seated in
the Luer fitting.
3. Rinse 10-15 liters of water through the filter to
drain prior to using the product water.
MAINTENANCE AND SERVICING
Note
Used cartridges may be recycled. See
P.U.R.E. information packed with new
cartridges.
Caution
Do not overtighten the 0.2 micron filter
assembly onto the Luer fitting or use
excessive force in seating it. The filter
and/or Luer fitting can be damaged by
overtightening or excessive force.
Note
If a newly installed 0.2 micron filter
clogs rapidly after installation, the
EASYpure UV/UF may need to be
sanitized to remove bacterial contami-
nants. See System Sanitization.