Specifications
21
Installing Float or Pressure
Switch
Accessories D0606 (float switch) and D2706 (pressure
switch) are designed to protect the EASYpure UV/UF
pump by alerting the EASYpure UV/UF of an inadequate
feedwater condition so that the pump can be shut down.
Use the following instructions for installation.
1. Disconnect the unit from the electrical power.
2. If using D0606 float switch, follow the installa-
tion instructions included with the float switch
for installation to tank.
3. If using D2706 low pressure switch, install the
PVC tee (supplied with D2706) in incoming
water line. Screw the switch into the top of the
tee, then connect the inlet tubing to the
EASYpure UV/UF with the remaining opening.
4. Route cable from float or low pressure switch
to the rear of the EASYpure UV/UF.
5. Remove jumper plug and save for future use.
6. Plug the cable into the jumper plug outlet.
7. Reconnect electrical power.
Figure 11: Float and Pressure Switch Installation
Warning
Disconnect from the power supply prior
to maintenance and servicing.