Specifications

20
1. Press the START/STOP button to start. The
EASYpure UV/UF’s pump will begin to run and
the Purity meter will display the resistivity of the
water in megohm-cm.
2. Allow the water’s resistivity to rise to the desired
purity before drawing off water.
3. The system should be left on during the work
day. (See use of Standby mode.)
Water Draw Off
1. Remove the protective cap from the filter bell.
2. Depress the draw-off lever.
3. When draw off is complete, lift the draw-off lever
and replace the protective cap on the filter bell.
Use of Standby Mode
At the end of the work day, press the Standby switch on the
front of the EASYpure UV/UF to place the unit in Standby
mode for the night.
Normal Operation
Note
On initial start-up, the purity meter may
display “ERR”. This is caused by air in
the cell and should be replaced by a
resistivity reading almost immediately.
If “ERR” does not go out after the
pump has run for a minute or if it
appears any time while the EASYpure
UV/UF is in operation, refer to the
Problem Solving section of this
manual.
Note
For low organic and pyrogen applica-
tions, draw off 50 to 100 ml of water
from system and discard prior to
drawing water for each use.
Note
Do not turn off the EASYpure UV/UF
during non-work hours. Doing so will
allow bacterial growth and other
contamination of the water in the
system. As a result, the system will
require a lengthy rinse-up period at the
beginning of the work day to achieve
high-quality product water.