10. Press the “Flush” button twice to begin an
extended flush. When water begins exiting the
system, close the draw-off valve.
11. When the extended flush ends, press the
START/STOP button to stop and depressur-
ize the system by opening the draw-off
valve and allowing water to drain from the
unit until draining ceases. Close the draw-
off valve. Open the cartridge access door and
repeat steps 2 - 6 with the EASYpure High
Purity/Low TOC (Catalog No. D50229)
cartridge, placing it in the left-hand position.
Do not install the 0.2 micron filter and bell
assembly at this time.
12. Close the cartridge access door. Press the
“Flush” button twice again to begin another
extended flush. When the extended flush
ends, open the draw-off valve and press the
START/STOP button. Rinse 10-15 liters of
water through the draw-off valve to drain.
Close the draw-off valve.
13. Remove a new 0.2 micron filter and bell
assembly from its bag and insert it into the
Luer fitting on the draw-off valve. Gently turn it
clockwise until it is fully seated in the Luer
fitting.
14. Remove the protective cap from the filter bell.
Open the draw-off valve and flush 10-15 liters
of water through the 0.2 micron filter.
INITIAL OPERATION
Figure 10: Proper Cartridge Position and Order
Pretreatment Cartridge
D50230
ULTRApure Cartridge
D50233
High Purity/Low TOC Cartridge
D50229
Warning
Use properly grounded electrical outlet
of correct voltage and current handling
capacity.
Caution
Do not allow EASYpure UV/UF to
operate unless water is available to
unit.
Note
For more demanding applications
where low TOC water is required, a
rinse of 15-20 liters through the car-
tridges and filter may be necessary.