Specifications

17
Initial Sanitization
Your EASYpure UV/UF has been shipped with a sanitiza-
tion cartridge (Catalog Number D50245) and two empty
cartridges to allow you to sanitize your EASYpure UV/UF
prior to using it for the first time. To sanitize your
EASYpure UV/UF:
1. Open the cartridge access door. Install the
supplied D50245 sanitization cartridge in the
empty cartridge position. Press the upper end
cap into the upper right position until it bottoms
out. Then, lower the cartridge and insert the
lower end cap into the lower socket until it is
firmly seated. The empty cartridges have been
installed for you at the factory. Close and latch
the cartridge access door.
2. Insert the quick disconnect insert on the
feedwater line into the feedwater inlet quick
disconnect on the back of the EASYpure UV/
UF.
3. Plug the power cord into the unit’s power entry
module and plug into a live outlet. Place the
main power switch to the on position. Place a
suitable container under the outlet valve.
Depress the draw-off lever to open the outlet
valve. Press the “Flush” button twice to begin
an extended flush.
4. Close the draw-off valve when solution exits
the system.
5. When the extended flush stops (5 minutes),
open the draw-off valve. Press the START/
STOP button to start. Draw off 2.0 liters of
solution through the draw-off valve. Close the
draw-off valve. Discard the solution.
6. Recirculate the disinfectant solution for
approximately 30-45 minutes.
7. Press the START/STOP button to stop.
Initial Operation
Warning
Avoid splashing disinfecting solutions
on clothing or skin.
Ensure all piping connections are tight
to avoid chemical leakage.
Ensure adequate ventilation.
Carefully follow manufacturer’s safety
instructions on labels of chemical
containers and material safety data
sheets.
This device is to be used with water
feeds only.
Sanitizing/cleaning agents must be
used in compliance with instructions in
this manual. Failure to comply with the
above could result in explosion and
personal injury.
Use of a properly grounded electrical
outlet of correct voltage and current
handling capacity.
Figure 9: Preparing To Sanitize
Sanitization Cartridge
D50245
Empty Cartridges