Specifications

13
Reject Water Connection
When the EASYpure UV/UF flushes its membrane, the
water used is sent to drain through this connection. To
install:
1. Locate the reject water tubing. This is the
piece of 1/4” O.D. tubing that has an adapter
on one end and no connector on the other
end. Remove the nut, grab ring, backup ring
and o-ring from the reject water fitting located
at the lower right of the unit. (See Figure 3 for
part identification and Figure 8 for fitting
location.)
2. Follow instructions under Tubing Adapter
Installation and connect the reject water
tubing to the reject connection on the
EASYpure UV/UF.
3. Route the other end of the reject water tubing
(the end with the adapter installed) to an
atmospherically vented drain. Ensure there
are no kinks in the tubing and that it proceeds
in a downward plane. If possible, use the
tubing adapter to permanently install the reject
water tubing in the drain.
INSTALLATION
Figure 7: Orientation of Swivel Base
Figure 8: EASYpure UV/UF Connections
Door Latch
Pump Interlock
Plug
Main Power
Switch
Power Entry
Module
Feedwater Inlet
Connection
Reject Water
Connection