you. Viewed this way, the product connector is
the one on the left. The reject connector is the
one on the right. (See Figure 5) Remove the nut,
stainless steel grab-ring, backup ring and o-ring
from each of the ultrafilter connections and
discard.
5. The three pieces of tubing that you disconnected
from the “T” in step 2 are the Feed, Reject and
Product lines for the ultrafilter. The Reject and
Product tubing nuts are labeled; the unlabeled
one is the Feed line. Connect the ultrafilter Feed,
Reject and Product lines to the appropriate
connectors on the ultrafilter, identified in step 4.
Ensure that the stainless-steel grab ring, backup
ring and o-ring from each connector remain on
each piece of tubing. Tighten the connectors by
hand; do not use a wrench.
6. Press the ultrafilter into the ultrafilter securing
clips inside the EASYpure UV/UF case. Ensure
that you maintain the proper orientation of the
ultrafilter during installation, with the feed con-
nection on the bottom and the reject and product
connections on the top.
INSTALLATION
Figure 5: Ultrafilter Connections Figure 6: Ultrafilter Orientation and UV Chamber Location
Product
Reject
Feed
Black Plasitc Cap on
UV Chamber
Proper Installation of
Ultrafilter