Instruction manual
9-4 EASYpure II RF Thermo Scientific
Section 9
Maintenance and Servicing
Membrane Switch Lead
Remove Front Cover
and Screws
Remove Top
Cover and Screws
Meter Board
Display Board
Disconnect Cell
Lead From Here
Disconnect Membrane
Switch Lead From Here
EMI Filter
EMI/RFI
Suppression Filter
(one loop around filter)
Cell
Bushing
O-ring
Figure 9-1. Disconnecting the Resistivity Cell
7. Disconnect the cell lead from the meter board, remove the EMI/RFI
suppression filter and gently pull the cable out of the EASYpure II RF
frame. Note orientation.
8. Unscrew and remove the cell.
Caution The cell electrodes are etched to improve wetting characteristics.
Do not mechanically abrade or damage this surface (i.e. do not clean with
a wire brush, sandpaper, etc.).
Do not immerse the entire cell assembly in cleaning solution, only the
electrode portion.
s
Cleaning Resistivity Cell
(continued)