Section 9 Maintenance and Servicing
Frequency of cleaning is difficult to determine because of variability in
feedwater and usage. Cleaning is necessary when residual deposits are
evident inside the feedwater reservoir or if a new 0.2 micron final filter
clogs rapidly after installation even though the cartridges were properly
rinsed before the 0.2 micron final filter was installed. To sanitize the
EASYpure II RF, remove the 0.2 micron final filter and attach the hose
barb that came installed on the unit. The purification cartridges must be
removed and replaced with empty cartridges. These must be ordered
separately. Contact Technical Services to order.
Warning Disconnect from the power supply prior to maintenance and
servicing.
Do not disassemble water lines or remove cartridges where spilled water
could contact equipment that requires electrical service. Electrical shock
hazard could result.
Refer servicing to qualified personnel.
s
Note Drain plug is not attached to unit; use care when removing it over an
open drain to avoid dropping it into the drain.
s
Note The cartridges will still contain water when removed. Therefore, you
will want to have a sink, bucket or other waterproof container available to
place them in after removal.
s
Note A small amount of water will drain from the cartridge when it is
disconnected from the lower socket. Plug the cartridges lower opening
with your finger to minimize water spillage while you finish removing the
cartridge.
s
Note Used cartridges may be recycled. See P.U.R.E. information packed
with new cartridges.
s
EASYpure II RF 9-1Thermo Scientific