Instruction manual

5-2 EASYpure II RF Thermo Scientific
Section 5
Installation
Refer to Figure 5-3 for the procedure below.
1. Open cartridge access door in the rear of the unit by pushing the door
latch down.
2. Remove a new Pretreatment cartridge from its plastic bag.
3. Wet the o-rings with water on both end caps.
4. Press the upper end cap into the pretreatment position until it bottoms
out.
5. Lower the cartridge and insert the lower end cap \into the lower socket
until it is firmly seated.
6. Repeat Steps 2 - 5 with the Ultrapure cartridge and then the EASYpure
High Purity/Low TOC or Ultrapure cartridge, placing them in
positions 2 and 3.
7. Close cartridge access door. This serves to verify the cartridges have
been properly seated.
Note The upper end cap is the one with the right-angle turn and the two
flanges. The lower end cap extends straight out from the cartridge.
Note The two flanges on the end cap should be able to slide down on each
side of the keyway wall.
Cartridge Installation
Figure 5-3. Cartridge Installation