Instruction manual

EASYpure II RF 3-1Thermo Scientific
Section 3 General Specifications
Dimension Requirements
12" W X 19" D X 18 1/8" H (30.5 cm X 48.3 cm X 46.0 cm).
Clearance Requirements
Sides - 4" (10.1 cm) minimum to allow air flow
Above - 12" (30.5 cm) minimum for reservoir replenishment
Cartridge replacement requires that the back of the unit be accessible to
open the cartridge access door (total depth, unit + open door, = 34") (86.4
cm).
Storage Reservoir Capacity - Approximately 6.5 liters usable, 7.0 liters total
Electrical Requirements
The EASYpure II RF is equipped with two power cords and corresponding
fuses taped to each power cord to be plugged into a grounded electrical
outlet of the appropriate voltage.
Model D7031: 100-240 VAC +5% -10%, 47-63 Hz.
Feedwater Requirements
The EASYpure II RF requires water pretreated by either distillation,
deionization or reverse osmosis.
TOC . . . . . . . . . . . . . . . . . . . . .Less than 1.0 ppm
Turbidity . . . . . . . . . . . . . . . .Less than 1.0 N.T.U.
Temperature . . . . . . . .40°F - 104°F (4.4°C - 40°C)
Silica . . . . . . . . . . . . . . . . . . . . . . .Less than 1 ppm
TDS (CaCO3)
RO - Less than 50 ppm
DI - Less than 100 ppm
Distilled - Less than 2 ppm
Product Water Quality
Resistivity . . . . . . . . . . . . . . . . . . . . .ASTM Type I
TOC . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15 ppb
Flow Rate . . . . . . .>
0.8 LPM with a new final filter