Specifications

8. Repeat steps 3-7 with the EASYpure Ultrapure
(Catalog No. D50233) cartridge, placing it in
position 2 and either the Ultrapure or the
EASYpure High Purity/Low TOC (Catalog No.
D50229) cartridge in position 3. Be sure to
remove and replace one cartridge at a time.
9. Close the cartridge access door and reattach
feedwater.
10. Remove the old 0.2 micron final filter(s) from the
draw-off valve and remote dispenser (if avail-
able.)
11. Install the hose barb into the fitting in the draw-
off valve.
12. Turn unit on and from “IdL” press the START
button to begin unit operation.
13. Rinse approximately 5 liters of water through the
draw-off valve or remote dispenser (if available)
to drain
.
14. Close the draw-off valve.
15. Sanitize the EASYpure II LF. Refer to the
System Sanitization section of this manual.
0.2 Micron Final Filter Replacement
Replace the 0.2 micron final filter whenever any of the fol-
lowing conditions occur: the product water flow rate is
reduced or you experience bacteria break through, when
cartridges are replaced, or when system is sanitized. The
0.2 micron final filter is shipped assembled with a bell. To
replace the 0.2 micron final filter assembly on draw-off
valve or remote dispenser:
1. Remove the old 0.2 micron final filter assembly
by turning it to the left to unscrew it from the
draw-off valve.
2. Remove the new 0.2 micron final filter assem-
bly from its bag and insert it into the draw-off
valve. Gently tighten, turning the filter to the
right.
3. Rinse 5-10 liters of water through the filter to
drain prior to using the product water.
MAINTENANCE AND SERVICING
Warning
Depressurize system prior to removing
cover.
Note
It is suggested that Teflon
®
tape be
applied to the threads of the 0.2 micron
final filter to ensure a tight seal.
27