Specifications

Cleaning the Resistivity Cell
(See Fig. 11 & 12)
1. Turn off the EASYpure II LF and disconnect it
from the power supply. Remove the power
cord.
2. Disconnect incoming water line at quick dis-
connect. Depressurize the system by opening
the optional remote dispenser or unit draw-off
valve, allowing water to drain until no more
flows from the valve(s).
3. Remove the screws securing the EASYpure II
LF top cover.
MAINTENANCE AND SERVICING
Warning
Depressurize system prior to opening
cartridge access door.
Note
Ensure you have an o-ring (part #
GSX29) available prior to cleaning
resistivity cell.
Disconnect membrane
switch lead from here
Disconnect cell
lead from here
EMI Filter
Membrane Switch Lead
Remove Top Cover
and Screws
Remove Front Cover
and Screws
Display Board
Meter Board
Check Valve
Figure 11: Disconnecting Resistivity Cell
24