Specifications

22
Sanitization
Your EASYpure II LF has been shipped with a sanitization
syringe (Cat. No. CMX25). When the system is new, or
whenever new cartridges are installed, the system should
be sanitized. Sanitize your EASYpure II LF as follows:
1. If the 0.2 micron final filter is installed, remove it
and attach the hose barb that came installed on
the unit.
2. Turn the system power off.
3. Disconnect the feed water supply at the inlet
quick disconnect connection.
4. Ensure that the remote dispenser (if applicable)
is open and pointed to drain. If no remote dis-
penser is present, ensure that the unit draw-off
valve is open and a container is placed under
the draw-off valve.
5. Remove the sanitization syringe from wrapper.
6. Open right side door of unit and remove the luer
cap on the sanitization block, located on the
inside left of the cartridge area.
7. Remove the syringe cap and attach syringe to
the luer fitting on the sanitization block.
8. Slowly inject the solution into the system. Some
water will be displaced from the remote dis-
penser or unit draw-off valve. Avoid injecting air
into the system.
9. Remove the syringe and replace the luer cap on
the sanitization block.
10. Close the remote dispenser or unit draw-off
valve.
11. Reconnect the feed water supply removed in
step 3.
12. Allow the system to stand for at least 1 hour.
13. If using an optional remote dispenser, turn sys-
tem power on and from “IdL”, start the system
by pressing the START switch and allow to run
Maintenance and Servicing
Figure 10: Sanitization
Warning
Disconnect from the power supply prior
to maintenance and servicing.
Do not disassemble water lines or re-
move cartridges where spilled water
could contact equipment that requires
electrical service. Electrical shock haz-
ard could result.
Refer servicing to qualified personnel.