Specifications

Power Connections
(See Fig. 8)
1. The power connection is located on the left side
of the unit, in the upper back corner.
2. Determine which power cord you need (this will
be based on your country and outlets available
in your lab). Both North American and
European power cords are provided with the
unit.
3. Remove the fuse holder, install the fuses includ-
ed with the power cord to be used, and reinstall
holder.
4. Verify power switch is turned off and attach
receptacle end of power cord into the power
socket.
5. Plug other end of power cord into facility power.
Air Purge
1. If the optional remote dispenser is connected to
the EASYpure II LF, place the draw-off trigger
into the locked position by depressing the button
on the right side of the dispenser at the same
time as the trigger is depressed.
OR
If there is no remote dispenser being utilized,
ensure that the hose barb is attached to the
draw-off valve fitting and depress the draw-off
valve lever to open the draw-off valve.
2. Place a suitable container under the opened
valve to collect water while purging air.
3. Connect the water supply to the EASYpure II LF
and open the customer supplied shut-off valve.
4. Turn the power switch on.
INSTALLATION
Warning
To avoid electrical shock, always use a
properly grounded electrical outlet of
correct voltage and current handling
capacity.
17
Caution
Do not allow EASYpure II LF to operate
unless water is available to unit.