Specifications
24
Problem Possible Causes Solutions
0.2 micron final filter clogs EASYpure LF contaminated Sanitize EASYpure LF according
rapidly after replacement. with bacteria. to the instructions in System
Sanitization. Replace the 0.2 filter
assembly as indicated in the 0.2
Micron Filter Replacement section.
Short cartridge life. Cartridges being used are beyond Check the expiration date. Cartridges
expiration date begin to lose capacity after being
stored two years from the date of
manufacture. Replace the cartridges
with unexpired ones.
Change in feedwater characteristics. If a Barnstead ROpure is the
feedwater source, check that the
membrane is functioning properly.
If a Barnstead Still is the feedwater
source, ensure that the distillate
temperature does not exceed 120°F
when added to the EASYpure LF
feedwater reservoir.
If feedwater is from a central water
purification system, verify water
quality and proper functioning of the
system.
Water leakage inside Cartridge connecting tubing not Press connecting tubing firmly into
EASYpure LF. fully seated into sockets. sockets.
Missing or defective cartridge Install or replace cartridge
O-rings. O-rings.
TROUBLESHOOTING