Specifications
22
14. Reinstall the cell into the cell well and hand tighten.
Reroute the cable up through the housing and recon-
nect.
15. Reinstall the screw holding the cell-cable retaining
clip. Reinstall membrane switch lead. Replace the
front cover. Retape the Luer fitting with new Teflon
tape and reinstall fitting. Replace the top cover.
16. Reconnect the power cord and the pump interlock
plug to the unit, and reconnect the unit to the power
supply. Connect incoming water line at quick discon-
nect.
Shutdown
If the EASYpure LF is to be shut down for an extended pe-
riod of time, the unit should be completely drained and the
cartridges removed to prevent the growth of bacteria.
If the system has remained inactive and full of water for
more than 96 hours, the unit should be drained, sanitized
and new cartridges installed prior to use.