Specifications

7
Dimensions and Clearance Requirements
EASYpure II UV/UF dimensions: 12 " W X 17 1/8" H X 19" D (30.5 cm x 43.5 cm x 48.3 cm).
Clearances:
Right side - 11" (27.9 cm) minimum for cartridge removal.
Left side - 12" (30.5 cm) minimum.
Above - 12" (30.5 cm) minimum for removal of the top cover and UV lamp replacement.
Electrical Requirements
The EASYpure II UV/UF is equipped with two power cords and corresponding fuses taped to each power cord
to be plugged into a grounded electrical outlet of the appropriate voltage.
Model D8611 - 100-240 VAC +5%, -10% at 47-63 Hz.
Feedwater and Drain Requirements
The EASYpure II UV/UF requires water pretreated by either distillation, deionization or reverse osmosis
meeting the following criteria:
TOC: Less than 1.0 ppm.
Turbidity: 1.0 N.T.U. maximum
Temperature: 40°F - 104°F (4.4°C - 40°C)
TDS:
RO - Less than 50 ppm as CaCO
3
DI - Less than 10 ppm as CaCO
3
Distilled - Less than 2 ppm as CaCO
3
Pressure Range: Gravity feed to 100 psig (7 kg/cm
2
) maximum.
Flow Rate: Greater than 0.26 gpm (1.0 lpm) at positive pressure.
Feedwater Source: to connect 1/4” NPT fitting with 10’ of 3/8” O.D. tubing provided
Atmospherically Vented Drain: 1/4” O.D. x 6’ long tubing provided
Product Water
Quality
Resistivity (Conductivity less than .055 µs/cm): Greater than 18.0 µ/cm
TOC: 1-5 ppb
Endotoxin: Less than 0.001 EU/mL
Flow rate
0.29 gpm (1.1 lpm) maximum at minimum inlet feedwater pressure of 30 psig (2.1 kg/cm
2
) and with a new
D3750 final filter.
Bacteria: Less than 1 CFU/mL
General Specifications