Specifications
34
Fuse Replacement (See Fig. 11)
1. Turn off the EASYpure II UV/UF and disconnect
it from the power supply. Remove the power
cord.
2. Pull out the fuse drawer located in the power
entry module.
3. Remove old fuses and replace with fuses of
the same type and rating. (See Consumables
section.)
4. Replace fuse drawer on left side of unit.
5. Reattach the power cord and reconnect the
unit to the power supply.
6. Operate normally.
Unit Shutdown
If the EASYpure II UV/UF will be inactive for a period up
to a month, place the unit in Standby. For periods of time
greater than a month, disconnect water and power
sources, drain unit, close customer supplied water valve,
and remove and discard cartridges, ultrafilter and final fil-
ter. See See Sanitization, Cartridge Replacement and
0.2 Micron Final Filter Replacement sections for reacti-
vation.
MAINTENANCE AND SERVICING
Warning
Replace fuses with those of the same
type and rating.