Specifications

Ultrafilter Replacement
1. Turn off the EASYpure II UV/UF and disconnect it
from the power supply. Remove the power cord.
2. Disconnect incoming water line at quick discon-
nect. Depressurize the system by opening the
optional remote dispenser or unit draw-off valve,
allowing water to drain until no more water flows
from the valve(s).
3. Follow the instructions for ultrafilter installation
found under Ultrafilter Installation, removing the
old ultrafilter the same way that is described in
Push-to-Connect Fitting Tubing Removal.
4. Sanitize the EASYpure II UV/UF. Refer to the
System Sanitization Procedure section of this
manual.
0.2 Micron Final Filter Replacement
Replace the 0.2 micron final filter whenever any of the fol-
lowing conditions occur: the product water flow rate is
reduced or you experience bacteria break through, when
cartridges are replaced, or when system is sanitized. The
0.2 micron final filter is shipped assembled with a bell. To
replace the 0.2 micron final filter assembly on draw-off
valve or remote dispenser:
1. Remove the old 0.2 micron final filter assembly
by turning it to the left to unscrew it from the
draw-off valve.
2. Remove the new 0.2 micron final filter assem-
bly from its bag and insert it into the draw-off
valve. Gently tighten, turning the filter to the
right.
3. Rinse 5-10 liters of water through the filter to
drain prior to using the product water.
MAINTENANCE AND SERVICING
Note
Some water will drain from the ultrafil-
ter when you disconnect it. Have a
container ready to hold the ultrafilter
and catch the water while it drains.
Water could also drain from the tubing
of the EASYpure II UV/UF onto your
bench or floor. To avoid slipping and
falling, clean up all spills.
33
Note
It is suggested that Teflon tape be
applied to the threads of the 0.2 micron
final filter to ensure a tight seal.