Specifications
32
2. Depressurize the system by disconnecting unit
feedwater, opening draw-off valve and allowing
water to drain from the unit until draining ceas-
es.
3. Remove the screws securing the EASYpure II
UV/UF top cover.
4. Remove the top cover by lifting straight up.
5. Locate the UV oxidation chamber and pull the
top black plastic cover off. Do not pull on the
cable.
6. While holding on to the lamp, remove the plug
from the lamp. Dispose of the UV lamp appro-
priately. The used lamp does contain mercury
vapor and should not be disposed of in the
trash. Recycle or dispose of the used lamp as
hazardous waste.
7. Remove the replacement lamp from its con-
tainer. DO NOT TOUCH THE GLASS POR-
TION OF THE LAMP. It is recommended that
lint free gloves be worn when handling the
lamp. The glass portion must be free of finger-
prints, perspiration, etc. Even a light coating of
perspiration will reduce the effectiveness of the
lamp.
8. Clean the lamp with isopropyl alcohol and a lint
free cloth.
9. Carefully insert and hold the UV lamp partially
into the UV chamber.
10. Connect the UV lamp to the receptacle in the
black plastic cover of the UV chamber.
Replace the black plastic cover on the UV
chamber.
11. Reinstall the EASYpure II UV/UF top cover.
12. Reattach the power cord and reconnect the
unit to the power supply.
MAINTENANCE AND SERVICING
Note
If UV lamp is not installed properly an
“Er.3” message will appear for 15 sec-
onds of every minute.
Note
The UV lamp contains mercury. If
broken or no longer needed, do not
dispose of the UV lamp in the trash.
Recycle or dispose of the UV lamp as
hazardous waste.