Specifications
Purity/Low TOC (Catalog No. D50229) cartridge,
placing it in position 3. Be sure to remove and
replace one cartridge at a time.
9. Close the cartridge access door and reattach
feedwater.
10. Remove the old 0.2 micron final filter(s) from the
draw-off valve and remote dispenser (if avail-
able.)
11. Install the hose barb into the fitting in the draw-
off valve.
12. Turn unit on and from “IdL” press the Flush but-
ton twice to begin a 5-minute flush.
13. When flushing ends and the unit returns to “IdL”,
open the draw-off valve and press the
START/STOP button to begin unit operation.
14. Rinse approximately 5 liters of water through the
draw-off valve or remote dispenser (if available)
to drain.
15. Close the draw-off valve.
16. Sanitize the EASYpure II UV/UF. Refer to the
System Sanitization Procedure section of this
manual.
UV Lamp Replacement (See Fig. 5)
Lamp life will vary according to the number of times the
EASYpure II UV/UF unit is turned on and off. If the
EASYpure II UV/UF is cycled between Run, Standby and
Idle Modes frequently during the work day, this will result
in a shorter lamp life. Therefore, it is recommended that
the EASYpure II UV/UF be left in the Run Mode during
normal working hours and in Standby Mode at night and
on weekends.
To replace the ultraviolet lamp:
1. Turn off the EASYpure II UV/UF and discon-
nect it from the power supply. Remove the
power cord.
MAINTENANCE AND SERVICING
Warning
This unit is equipped with an ultraviolet
lamp. Ultraviolet radiation is harmful to
the eyes and skin. Do not attempt to
observe the lamp directly.
Caution
Do not unscrew the metal end cap
cover of the UV chamber, as this will
loosen the water tight seal and may
damage the replacement lamp. Pull the
black plastic cover straight up.
Warning
Depressurize system prior to removing
cover.
31