Specifications

Remove Top
Cover and Screws
Daughter Board
Check Valve
Membrane Switch Lead
Meter Board
Remove Front Cover
and Screws
EMI Filter
Disconnect Membrane
Switch Lead From Here
Disconnect Cell
Lead From Here
Display Board
28
Cleaning the Resistivity Cell
(See Fig. 14)
1. Turn off the EASYpure II UV/UF and discon-
nect it from the power supply. Remove the
power cord.
2. Disconnect incoming water line at quick dis-
connect. Depressurize the system by opening
the optional remote dispenser or unit draw-off
valve, allowing water to drain until no more
flows from the valve(s).
3. Remove the screws securing the EASYpure II
UV/UF top cover.
MAINTENANCE AND SERVICING
Warning
Depressurize system prior to opening
cartridge access door.
Note
Ensure you have an o-ring (part #
GSX29) available prior to cleaning
resistivity cell.
Figure 14: Disconnecting Resistivity Cell