Specifications

27
14. Open the remote dispenser and point to drain.
If no remote dispenser is present, ensure the
unit draw-off valve is open and a container is
placed under the hose barb that was attached to
the draw-off valve fitting in step 1.
15. From “IdL” press the Flush switch twice to rinse
the sanitant from the system during a 5-minute
flush.
16. After the 5-minute flush is completed, close
remote or unit draw-off valve.
17. Remove the hose barb attached in Step 1.
Keep this hose barb for future use.
18. Remove a new 0.2 micron final filter and bell
assembly from its bag and insert it into the fitting
in the draw-off valve and remote dispenser (if
available.) Gently turn it clockwise until it is fully
seated.
19. Remove the protective cap from the filter bell.
Open the draw-off valve or remote dispenser
and flush 10-15 liters of water through the 0.2
micron final filter.
20. The sanitization is now complete.
MAINTENANCE AND SERVICING
Warning
Avoid splashing disinfecting solutions
on clothing or skin.
Ensure all piping connections are tight
to avoid chemical leakage.
Ensure adequate ventilation.
Carefully follow manufacturer’s safety
instructions on labels of chemical con-
tainers and material safety data
sheets.
Note
It is suggested that Teflon
®
tape be
applied to the threads of the 0.2 micron
final filter to ensure a tight seal.
Teflon
®
is a registered trademark of DuPont