Specifications
Standby
In Standby (“SbY”) Mode, the pump will restart for ten min-
utes out of every hour (i.e 50 minutes off, 10 minutes on),
during which time the display reads “rEC” which indicates
recirculation. Every fourth time that the pump turns on, the
UV lamp will also turn on for 10 minutes. This will allow the
unit to produce high quality water quickly upon being placed
in the Run Mode and prolong the life of the UV lamp. It is
recommended that the EASYpure II UV/UF be placed in the
Standby Mode during non-work hours. At the end of the
work day, press the STANDBY switch to place the unit in
Standby mode. “SbY” will appear on the display.
FLUSH
Manual Flush
Pressing the Flush switch once will initiate a one minute
flush of the ultrafilter. During this one minute flush, water is
sent to drain through the drain tubing and the display will
show FL1. Pressing the Flush switch twice will initiate a five
minute flush of the ultrafilter. During this five minute flush
the display will count down FL5, FL4, FL3, etc. After the
timed flush has ended, the EASYpure II UV/UF will return to
the mode (“IdL”, Run or Standby) it was in when you initiat-
ed the flush. We recommend that you initiate a one minute
manual flush when starting up the EASYpure II UV/UF prior
to your day’s work.
Automatic Flush
In both the Run and Standby Modes, the EASYpure II
UV/UF will initiate a one minute automatic flush of the ultra-
filter once every twenty-four hours of operation. This pre-
vents bacterial buildup on the ultrafilter. A microcontroller
timer in the EASYpure II UV/UF counts the hours power is
applied to the unit. When the timer reaches twenty-four
hours, the EASYpure II UV/UF initiates a one minute flush if
in the Run Mode. If the unit was in Standby, the automatic
flush will occur immediately after the START/STOP switch is
pressed and Run Mode entered. The timer will reset to
zero when the one minute automatic flush is completed.
Idle Mode (“IdL”)
“IdL” indicates the unit is powered and waiting to be placed
in Run, Standby or Flush Mode.
24
C
ONTROLS AND NORMAL
OPERATION
Note
If the unit is in Standby and power to
the unit is turned off or lost, the unit
will return to Standby once power is
restored.