Specifications
Power Switch
The power switch on the EASYpure II UV/UF is located
on the left side of the unit, directly above the power cord
receptacle.
Control Panel
The EASYpure II UV/UF control panel incorporates three
switches and a digital display.
Switches (See Figure 12)
When the main power switch is on, the three switches on
the control panel function as follows:
1. START/STOP: When the unit is in the Idle Mode
(“IdL”), pressing the START/STOP switch will
put the unit in the Run Mode, turning the pump
and UV lamp on. When the EASYpure II UV/UF
is in the Standby (“SbY”) Mode, pressing the
START/STOP switch will return the unit to Run
Mode. When the unit is in the Run Mode,
pressing the START/STOP switch will put the
unit into the “IdL” Mode.
2. STANDBY: Pressing the STANDBY switch will
place the unit in “SbY” Mode from either the
Run Mode or the “IdL” Mode. Pressing the
STANDBY switch while the unit is in “SbY”
mode has no further effect.
3. FLUSH: Manual Flush: Pressing the Flush
switch once will initiate a one minute flush of the
ultrafilter and the display will show FL1.
Pressing the Flush switch twice will initiate a five
minute flush of the ultrafilter. During this five
minute flush, the display will count down FL5,
FL4, FL3, etc. After the timed flush has ended,
the EASYpure II UV/UF will return to the mode
(“IdL”, Run or Standby) it was in when you initi-
ated the flush.
Automatic Flush: In both the Run and Standby
Modes, the EASYpure II UV/UF will initiate a
one minute automatic flush of the ultrafilter once
every twenty-four hours of operation.
22
Controls and Normal Operation
Figure 12: Control Panel
FLUSH
START/STOP
STANDBY
Note
When the unit is first put in the Run
Mode, the display will show 10.0 for a
few seconds. This is an arbitrary
number that indicates the unit is run-
ning. Any number that appears after
10.0 indicates purity.
Note
On initial power-up, the display will run
the following sequence:
- Model type is scrolled Ut-UF
(Note: UV is displayed as Ut),
- The display’s LEDs will light up,
- Followed by the unit software
revision,
- Finally, IdL will be displayed.