Specifications
2. Install the tubing adapter into the atmospherical-
ly vented drain and route the tubing to the
EASYpure II UV/UF drain connector, ensuring
that there are no kinks
3. Take the tubing end that has no fitting, wet the
tube end with water and insert the tube straight
into the unit drain connector until it bottoms out.
Refer to section Push-to-Connect Fitting
Tubing Installation for more explicit details.
Feed Water
1. Locate the length of 3/8” O.D. tubing provided
with a quick disconnect insert on one end and
a 3/8” O.D. X 1/4 NPT tubing adapter on the
other.
2. Install the tubing adapter into your incoming
water line. We recommend a customer sup-
plied shut off valve be installed in your feed
water line. Do not connect the feedwater to
your EASYpure II UV/UF until you are told to
do so. (See Air Purge section.)
Power Connections
(See Fig. 11)
1. The power connection is located on the left side
of the unit, in the upper back corner.
2. Determine which power cord you need (this will
be based on your country and outlets available
in your lab). Both North American and
European power cords are provided with the
unit.
3. Remove the fuse holder, install the fuses includ-
ed with the power cord to be used, and reinstall
holder.
4. Verify power switch is turned off and attach
receptacle end of power cord into the power
socket.
5. Plug other end of power cord into facility power.
20
I
NSTALLATION
Caution
Do not connect feed water until direct-
ed to do so during the Air Purge sec-
tion.
Warning
To avoid electrical shock, always use a
properly grounded electrical outlet of
correct voltage and current handling
capacity.