Specifications
Remote Dispenser (Optional Accessory
Part No. D11981)
The remote dispenser is best installed after the EASYpure
II UV/UF system has been either bench or wall mounted.
The remote dispenser can be located up to 6’ away from
the unit. Please refer to the remote dispenser operating
and installation instructions for additional information.
Tube Connections
Prepare the EASYpure II UV/UF system for attaching
the remote dispenser as follows:
1. Make sure unit is depressurized, water supply is
disconnected and unit power is disconnected.
2. Using the tube removal tool supplied with dis-
penser, disconnect “U” shaped tube from left
rear of unit as shown. Press in on collet of bulk-
heads with tool while pulling on tube. Work back
and forth from each bulkhead until tube is free.
Refer to section Push-to-Connect Fitting
Tubing Removal for further details.
3. Mark or note orientation of tubes as “Remote 1”
and “Remote 2”. The recirculating and feed
lines of the dispenser are NOT interchangeable,
so orientation of the two connections is impor-
tant. Remove and discard fittings on the end of
the dispenser tubes.
4. Press tube labeled “Remote 1” into bulkhead
labeled “1” as shown. Refer to section Push-to-
Connect Fitting Tubing Installation for further
details.
5. Press tube labeled “Remote 2” into bulkhead
labeled “2” as shown.
6. The remote dispenser is now properly attached
to the EASYpure II UV/UF.
7. Do not install the 0.2 micron final filter until after
the air purge and sanitization procedures have
been completed.
18
I
NSTALLATION
Note
The recirculating and feed lines of the
dispenser are NOT interchangeable,
so orientation of the two connections
is important.
Warning
To avoid electrical shock, always dis-
connect from power supply before
maintenance and servicing. Refer
servicing to qualified personnel.
Figure 9: Tube Connections
Note
Save U-shaped tube for future use.
Bulkhead 1
Bulkhead 2
Tube removal tool
supplied with
dispenser
Fittings to be
removed and
discarded