Specifications
Locating and Mounting Unit
Bench Mounting
Place the EASYpure II UV/UF on a flat, level bench top
that is accessible to water, electricity and an atmo-
spherically vented drain, and that is convenient to your
work area, noting the Clearance Requirements.
Wall Mounting
Install the EASYpure II UV/UF on a wall in a convenient
location that is accessible to water, an atmospherically
vented drain and electricity.
1. Locate the wall bracket packed separately from
the unit.
2. Using the wall bracket as a template, locate and
drill the mounting holes in the wall. A minimum
of four (customer supplied) fasteners will be
required — two on the top and two on the bot-
tom.
3. Attach the wall bracket to the wall using the cus-
tomer supplied fasteners.
4. Remove the locking screws on each side of the
wall bracket.
5. Pull the two locking slides on each side of the
wall bracket out as far as they will go.
6. Hang the unit on the wall bracket by sliding the
mounting pins into the wall bracket slots.
7. Push the locking slides on each side of the wall
bracket in as far as they will go.
8. Replace the locking screws.
The optional remote dispenser is now ready to be
attached to the EASYpure II UV/UF.
17
I
NSTALLATION
Caution
Wall composition, condition and con-
struction, as well as fastener type,
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs.
Inadequate support and/or fasteners
may result in damage to mounting sur-
face and/or equipment. If you are un-
sure of mounting surface composition,
condition and construction or correct
fasteners, consult your building mainte-
nance group or contractor.
Note
Please refer to the General
Specifications section for clearance
requirements.
Note
It is important that the storage reser-
voir outlet to the EASYpure II UV/UF,
be at the same level
or higher than
the inlet of the EASYpure II UV/UF
system.
Warning
Do not locate the EASYpure II UV/UF
directly over equipment that requires
electrical service. Routine maintenance
of this unit may involve water spillage
and subsequent electrical shock haz-
ard if improperly located.
Warning
Do not use in the presence of flam-
mable or combustible materials; fire
or explosion may result. This device
contains components which may
ignite such materials.
Warning
Do not use in the presence of highly
corrosive substances such as bleach
or acid baths; fire may result.