Note When the Routine Care light is illuminated, you will need to scroll
down the list of message errors to determine which routine alert error is
signaling.
s
Type II Water System 5-15Thermo Scientific
Section 5
Maintenance and Servicing
Routine Care Alerts
Message Possible Causes Action
Replace Carbon
Chlorine capacity of the carbonpre-
filter has been reached, or 1 year
service life has expired.
Replace Carbon Prefilter
(D502115) and reset carbon
prefilter timer.
Check Prefilter
System has sensed that the prefilter
is plugged.
Replace the RO Prefilter
(D502113) and reset prefilter
timer.
Replace Prefilter
1 year service life of the prefilter has
expired.
Replace the RO Prefilter
(D502113) and reset the pre-
filter timer.
Replace Membrane
1 year service life of the membrane(s)
has expired.
Replace the RO
membrane(s)(FLX1265X1) and
reset the membrane timer.
Replace Res Filter
6 month service life of vent filter has
expired.
Replace vent filter element
(25001) and reset the vent
filter timer.
Replace MPS Cart
1 year service life of the MPS
cartridge has expired.
Replace the MPS cartridge
(D502114) and reset the MPS
cartridge timer.
Replace DI Pack
1 year service life of the cartridge
pack has expired.
Sanitize the system, replace
the cartridge pack (D502137),
and reset the sanitization and
cartridge pack timers.
Replace System UV
6 month peak service life of the
system UV lamp has expired.
Replace the system UV lamp
(LMX13) and reset the system
UV lamp timer.
Replace Res UV
1 year peak service life of the system
UV lamp has expired.
Replace reservoir UV lamp
(LMX31) and reset reservoir
UV lamp timer.
Sanitize System
1 year system sanitization/cleaning
reminder has expired.
Clean the system and reset
the sanitization timer.