Instruction manual

9. Press ALT-F10 for configuration screen.
10. Enter into Line Settings screen by typing ALT-P.
11. Type in 11 <Enter>, sets parameters as 9600, 8, N, 1.
12. Type in 21 <Enter>, sets to (COM2:) or (Type in 20 <Enter>, sets to
(COM1:), depending on your computer.
13. Type in 24 <Enter>, saves settings.
14. Press ESC to exit setup menu.
15. Procomm is now ready to accept input from the Barnstead TII.
16. To begin downloading a file, Type ALT-F1, and give the file an
appropriate name when prompted.
The optional printer (part no. AY1137X1) provides a paper tape record for
future reference. During normal recirculation, the printer prints date, time,
purity and temperature readings once every minute.
1. Make sure the printer and Barnstead TII are turned OFF.
2. Connect the Barnstead TII to the printer via the RS232 port on the
back left side of the unit. Use the tan printer cable included with the
Epson printer. Plug the printer power cord into an electrical outlet.
3. Turn power on to the Barnstead TII and printer.
4. Press MENU and use the up or down arrows until display reads,
“SYSTEM OPTIONS.
5. Press OK.
6. From the “Printer Attached” display, press the up or down arrow for
“YES”.
7. Press OK.
8. Press the BACK arrow twice to return to the STANDBY mode.
9. During normal recirculation mode, a purity and temperature reading
will be printed once every 12 minutes.
Type II Water System 3-23Thermo Scientific
Section 3
Installation
Connecting and Starting
the Printer
Procomm (continued)