7. Follow the prompts on the screen to rinse membranes. The prompts
will ensure that you do not have a tank connected to the TII unit by
asking if the tank is empty and if the tubing from the tank goes to
drain. Answer yes to both questions. Once initiated, the rinse
procedure will last two hours, and will stop automatically.
Caution Do not allow the Barnstead TII to operate unless water is available
to the unit.
s
Warning Use a properly grounded electrical outlet of correct voltage and
current handling capacity.
s
After performing an RO membrane rinse, connect the TII to a storage
reservoir. Refer to the Barnstead TII Storage Reservoir manual.
Note Install reservoir at the same height or higher than the TII system.
Outlet of the reservoir must be at or above the inlet of the TII.
s
3-14 Type II Water System Thermo Scientific
Section 3
Installation
RO Membrane Rinse
(continued)
Reservoir Connection
Figure 3-19. Barnstead TII and Barnstead TII Reservoir System Set-Up