Instruction manual

Place the Barnstead TII on a bench top that is accessible to electricity and
an atmospherically vented drain.
Wall composition, condition and construction, as well as fastener type,
must be considered when mounting this unit. The mounting surface and
fasteners selected must be capable of supporting a minimum of 150 lbs.
(68kg). Inadequate support and/or fasteners may result in damage to
mounting surface and/or equipment. If you are unsure of mounting
surface composition, condition and construction or correct fasteners,
consult your building maintenance group or contractor. Refer to Section 2
Technical Specifications, for clearance requirements.
Warning Do not locate the Barnstead TII directly over equipment that
requires electrical service. Routine maintenance of this unit may involve
water spillage and subsequent electrical shock hazard if improperly located.
Install the Barnstead TII on a wall in a convenient location that is
accessible to an atmospherically vented drain and electricity.
1. Locate the wall bracket packed separately from the unit.
2. Using the wall bracket as a template, locate and drill the mounting
holes in the wall. A minimum of four (customer-supplied) fasteners
will be required — two on the top and two on the bottom.
3. Attach the wall bracket to the wall using customer-supplied fasteners.
4. Remove the locking screws on each side of the wall bracket.
5. Pull the two locking slides on each side of the wall bracket out as far as
they will go.
6. Hang the unit on the wall bracket by sliding the mounting pins into
the wall bracket slots.
7. Push the locking slides on each side of the wall bracket in as far as they
will go.
8. Replace the locking screws.
Note The outlet of Barnstead TII storage reservoir must be above or at the
same level as the inlet of the Barnstead TII.
s
Type II Water System 3-3Thermo Scientific
Section 3
Installation
Bench Mounting
Wall Mounting