Instruction manual

Type II Water System 2-3Thermo Scientific
Section 2
Product Information
TII Water
Specifications (cont.)
Table 2-4. DI Product Water Quality (dispensed from unit)
Table 2-5. Dimensions and Clearance Requirements
Table 2-6. Environmental Conditions
Resistivity ASTM Type II, >15 Megohm-cm
TOC <15 PPB
Flow Rate Up to 1 LPM with a new D3750 final filter
Bacteria Less than 0.01 CFU/ml
Volumetric Dispense Accuracy: ±5%, Repeatability: ±3%
Technical
Specifications
Barnstead TII Dimensions 13.5" W X 18.5" D X 19.5" H (34.3 cm X 47.0 cm X 49.5 cm)
Clearances Sides - 9" (22.9 cm) minimum to allow for servicing
Above -3" (7.6 cm) minimum for removal of top cover
Front - 4.75" (12.1 cm) minimum for opening front door
Electrical Requirements
Barnstead TII is equipped with four power cords and correspon-
ding fuses taped to each power cord to be plugged into a ground-
ed electrical outlet of the appropriate voltage.
All models: 90-240 VAC, +5% -10%, 47-63 Hz.
Operating 15°C -30°C; 20% -80% RH, non-condensing.
Installation Category II (over-voltage) in accordance with IEC 664.
Pollution Degree 2 in accordance with IEC 664.
Altitude limit 6562 ft (2,000 meters)
Storage -25°C -65°C; 10% to 85% relative humidity