Instruction manual
2-2 Type II Water System Thermo Scientific
Section 2
Product Information
Table 2-1. Feed Water Requirements
1
1500 ppm NaCl feed water solution @ 25°C (77°F), @ 4.5 bar (65 psig), 15%
recovery. Flow rate and recovery will decrease with lower water temperature and
pressure. We recommend the use of a hot water mixing valve before the RO system for
water cooler than 15°C (59°F). Feed water suitability must be determined by a water
analysis.
Table 2-2. RO System
Table 2-3. Membrane Performance
2
2
500 ppm NaCl feed water solution @ 25°C (77°F), @ 4.5 bar (65 psig), 15%
recovery. Flow rate and recovery will decrease with lower water temperature and
pressure. The use of a hot water mixing valve before the RO system is recommended for
water cooler than 15°C (59°F).
Types Tap
1
(Potable)
Turbidity 1.0 N.T.U. maximum
Pressure Range 2-6.9 bar (30-100psig)
Temperature Range 4-40°C (39-104°F)
Minimum Inlet Flow Requirements 100 lph
pH 3-10
TDS 600 (max. ppm as CaCO3)
Silt Density Index < 5%
Free Chlorine < 3 ppm
Langlier Saturation Index < 1
Iron (total as Fe) < 0.5 ppm
Silica < 30 ppm
TII Water
Specifications
Model RO Product Flow Rate Recovery
D14031 12 lph (3.2gph), ±15% 10-20%
D14041 24 lph (6.3gph), ±15% 20-40%
Rejection
Inorganic (minimum) >90%
Inorganic (typical) >95%
Particles >99%
Bacteria >99%
Organics (>300 MW) >99%